I have had problems with my Resmed S9 APAP staying at 20 cm for 2-3 hours at a time which wrecks me the next day.

This problem has been ongoing for many months and when I think I have fixed it it comes back

Using Sleepyhead I can see that during this spell of high pressure it looks like I am double inhaling.

Is this possible??

Views: 1234

Reply to This

Replies to This Discussion

I think its time to make a decision.

Either buy another make of APAP or get referred to another hospital which supplies a different make

Before I make a decision I am going to try something different.

I have noticed that when it goes up to 20 most times it has done so when the pressure has dropped to minimum which is 12 at present.

I will increase the minimum pressure to 14 so hopefully this will keep my airways open.

Might work

Don't rule out a fault with the machine or restricted airflow.

I would also not rule out that you might be one of the few that needs a different machine.

In any case, it should not go to 20 and sit there, I would get it checked or replaced.

If you get another machine and it does the same then the problem is with you rather than with the machine.

Have you contacted the hospital and complained about it?

I would stick with ResMed gear if I could, it is considered the best option by many.

However, I have used Resperonics as well and it worked fine, but the ResMed was a little quicker in reacting to events.

However, it looks like your machine is over reacting.  If you are awake and it still sits at 20 then complain it is faulty.

Insist on getting another machine.

Its me not the machine.

The S9 has been checked by Resmed and no fault was found but the circuit board was changed as a precaution.

I have used another S9 but had the same problem.

I have been told by the hospital that despite my AHI being under 1 most of the time my Apnea was getting worse so I needed the pressure to be at 20

I thought I had found the cause of the problem as I found that the Climateline hose had developed a kink where it was suspended by the hose lift. Buying a new hose seemed to have fixed the problem.

I was advised by Resmed that despite using a full face mask my nasal problems, Polyps and deviated septum, could be causing the problems

I have found a old compliance report from when I used a Remstar. and even though the maximum was set to 20 the pressure rarely went above 15.

So Im tempted to buy a machine ,either a Sleep cube or another Philips APAP 

Or the other option is to get referred to another local hospital which supplies Philips 



Sleep2snore said:

Don't rule out a fault with the machine or restricted airflow.

I would also not rule out that you might be one of the few that needs a different machine.

In any case, it should not go to 20 and sit there, I would get it checked or replaced.

If you get another machine and it does the same then the problem is with you rather than with the machine.

Have you contacted the hospital and complained about it?

I would stick with ResMed gear if I could, it is considered the best option by many.

However, I have used Resperonics as well and it worked fine, but the ResMed was a little quicker in reacting to events.

However, it looks like your machine is over reacting.  If you are awake and it still sits at 20 then complain it is faulty.

Insist on getting another machine.

I have finally found the answer 

I have increased the EPR to 3

This has kept my pressure below 14 for the last 3 days .

Im very annoyed that the hospital and Resmed had not increased this 

For some reason increasing the EPR has reduced the Flow Limitations which is the reverse of what should happen.

If it works .........

Hummmm!  That just shouldn't work!

There must be something unique to the way you breath when sleeping confusing the machine.

It is all a bit strange and a bit confusing!  The Resperonics machines are slightly slower to respond, so it may have been this and setting the EPR to 3 on the ResMed machine that might just be holding off the response for the ResMed machine.  It is pure speculation though.

I think the combination of nasal polyps ,deviated septum and COPD is too much for the Resmed to sort out.

The polyp in my left nostril does act like a one way valve and closes on exhale which may explain why the EPR increase helps. It looks like the only option as far as masks is to get a Oracle Yuck!!!!! 

I have also seen my GP who is going to refer me to a hospital that supplies Philips Dreamstations with humidifiers!!!!

Lastly having nasal surgery is out of the question as its too risky. As the ENT consultant said "I can still breath through my mouth"

That sounds like a get out for the ENT surgeon.

There are a lot of people gone through this op' and even bigger op's to the nose.

Maybe he is just not up to it?

Im on Warfarin for life because of leg DVT,s and Warfarin would be stopped 5 days before the operation. This could leave me at risk of another DVT.. Also the last time my Warfarin was stopped I has to self inject Heparin so my belly turned into a very bruised pin cushion.

Also I weigh 135Kg so a general anesthetic would be risky

As the consultant said " I can still breath through my mouth"

 



Sleep2snore said:

That sounds like a get out for the ENT surgeon.

There are a lot of people gone through this op' and even bigger op's to the nose.

Maybe he is just not up to it?

I have found out why setting the EPR to 3 resolves the problem with high pressure .

Setting the EPR to 3 smoothes out the waveform between exhale and inhale . I have read this about Remstar A flex but not Resmeds EPR.

I have Sleepyhead graphs showing that between Exhale and Inhale the waveform stays on the zero line which is not the case when the EPR is set to 2

I thought EPR was just a comfort feature

EPR just reduces the pressure when you breathe out, this makes it easier to breathe out when on higher pressure.

Current thought is that the higher setting the EPR is the higher pressure is required, however I have not found this, but it does affect my AHI, it goes up slightly when using a higher number of EPR!

Despite changing to a Airsense 10 Autoset Im still having the high pressures and the Sleepyhead waveforms still show that im double inhaling .

In really struggling now . Its taking me all day to recover from the battering the machine gives me at night .

Reply to Discussion

RSS

Sleep Apnoea Forum

New to the Sleep Apnoea Forum? 

1. Stop by our Sleep Apnoea Welcome Center to introduce yourself to the SleepGuide community.
2. Start a New Topic of Conversation.
3. Post your photos - of yourself, your old CPAP machine, your new CPAP machine, your pet, something about you!

Interested in advertising, have a problem or need to contact us? Click the Report an Issue page.

 

Latest Activity

Jonathan replied to Jonathan's discussion ResMed AirSense 10 getting noisy
"Just to tidy this up, I finally had an in-person review by my sleep clinic yesterday, who agreed the ResMed AirSense 10 was "wheezy" and replaced the machine."
Nov 9
Collette rose replied to Collette rose's discussion 1st time and I’ve failed the mask leak test
"thank you jonathan,i did get the it pack so tried the small and still struggled but i was sat up in bed so will try lying down great suggestion "
Nov 3
Jonathan replied to Collette rose's discussion 1st time and I’ve failed the mask leak test
"Hi My first thought is "don't panic"!  We're all a bit different, and there are many things that can be tweaked.  I took to CPAP therapy like a duck to water, but I've had friends who say they can't cope, but…"
Nov 2
Collette rose posted a discussion

1st time and I’ve failed the mask leak test

Hi everyone, today I received my rental kit - prisma smart max, plus a Philips dreamwear mask which I’ve been wearing unconnected to try and get used to it. I just did a mask fit test using a medium mask and got a constant red/orange leak warning…See More
Nov 2
Collette rose is now a member of Hope2SleepGuide
Oct 31
John posted a discussion
Oct 30
Profile IconBill Wright, Nicole Watson, TJ Bennett and 2 more joined Hope2SleepGuide
Oct 29
John replied to John's discussion evil mask noise
"Ha well it lasted a couple of days but started howling again. But it's a step in the right direction, just needs a bit more work."
Oct 24
Jonathan replied to John's discussion Has anyone been told root cause of their OSA?
"Nope, I just accept that some people's throats tend to close when they sleep, due to a loss of muscle tone during deeper sleep.  Back when I was diagnosed, the sleep lab just filmed you whilst you slept, stuck an O2 meter on your finger,…"
Oct 24
Jonathan replied to John's discussion evil mask noise
"Good to know you found a solution!"
Oct 24
John replied to John's discussion Has anyone been told root cause of their OSA?
"Did anyone tell you exactly what the anatomical issue is, Jonathan?"
Oct 21
John replied to Jonathan's discussion ResMed AirSense 10 getting noisy
"Can't say I know about this."
Oct 20
John replied to John's discussion evil mask noise
"So I tore some fabric from one of the filters and glued it over the holes - and it worked!It was dead quiet, and the air was still flowing freely.It was quiet all night then after a few hours a bit of whistling until I gave it a wipe.Vast…"
Oct 19
John replied to John's discussion evil mask noise
"Hi Jonathan, well see my nose is blocked a lot, so I mouth breath which really sucks a lot more moisture and makes my mouth dry. So I put humidity on maximum, which makes for condensation at the mask end, which becomes a demonic reed instrument.Now…"
Oct 17
Jonathan posted a discussion

ResMed AirSense 10 getting noisy

HelloI have an NHS issued ResMed AirSense 10 with heated humidifier.  In the last couple of weeks, it's developed an annoying noise, when I inhale. It bothers me a little and my partner a lot! I sometimes get problems if the water container in the…See More
Oct 16
Jonathan replied to John's discussion evil mask noise
"Hi My first question is why is condensation forming?  I very occasionally get some condensation, may be once a month, but I can quickly run a finger around the inside of the mask, and wipe it away.  I use a nasal mask, with the humidifier…"
Oct 16
John replied to John's discussion What's a good Apnea forum?
"11K+? Ok that's probably where I should look then :-)"
Oct 16
Kath Hope replied to John's discussion What's a good Apnea forum?
"Hi John. Thanks for the compliment as I do my best popping on here, but it's difficult due to us now supporting 25k+ patients at the charity. To be honest, the forum is quiet because most people seem to prefer Facebook and forums aren't as…"
Oct 16
John posted discussions
Oct 16
Jonathan replied to John's discussion Has anyone been told root cause of their OSA?
"There's some information here, but generally it's very common and likely due to our anatomy."
Oct 12

© 2024   Created by The SleepGuide Crew.   Powered by

Badges  |  Report an Issue  |  Terms of Service